Description
Specifications :
› InputWay : Floating and defibrillation protection
› Time constant : Time constant > 3.2s
› CMRR :> 60dB, > 100dB (Adding Filter)
› EMG interference filter : 35 Hz (-3dB)
› Recording way : Thermal printing system
› Specification of recording paper : 210mm (W) x 20m (L) high-speed thermal paper
› Paper speed : Auto record : 25mm/s, 50mm/s, error : ±5%
› Rhythm record : 25mm/s, 50mm/s, error : ±5%
› Manual record : 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 50mm/s, error : ±5%
› Sensitivity selections : 5, 10, 20, 40mm/mV, error : ± 5%. Standard sensitivity is 10mm/mV ± 2mm/mV.
› Auto record :Record setup according to auto record format and mode, automatically changing leads, measuring and analyzing.
› Rhythm record :Record setup according to rhythm record format and mode, automatically measuring and analyzing.
› Manual record :Record setup according to manual record format, manually changing leads.
› Measurement parameters : HR, P-R. interval, P Duration, QRS Duration, T Duration, Q-T interval, Q-Tc, P Axis, QRS Axis, T Axis, R(VS), S(VI), R(VS) + S(VI)
› Product safety type : Class I CF applied part; there is defibrillation and packing protection circuit in it.
› Enduring polarization voltage : ± 500mV
› Noise level :<15μVp-p
› Size : 334mm (L) x 320mm (W) x 85m (H)
› Power supply / Frequency : AC : 100 – 240V, 50/60Hz DC:14.8V, 3700mAh rechargeable lithium battery