Description
Technical specifications
-Defibrillator type Manual, Synchronized, A synchronized
-Out put wave form Biphasic technology
-Energy accuracy<+1%
-Energy selection 0, 20, 50, 100, 160, 250, 300, 360 joule (nominal at a resistance of 50)
-Charging time Maximum 8 seconds to 360 Joules
-Standard Adult/Pediatric Paddles Reusable external adult paddles (pediatric paddles integrated)
Power requirements
-AC Power supply110V, 60 Hz /220V,50 Hz
-Vehicle Voltage DC 12V
Internal Battery Backup
-Battery type Rechargeable lead-acid battery
-Charging time Time-Minimum of 4 hours for full charging
-Battery charging During AC Power Supply operation of the device.
-Charging /Low Battery Indication Red LED
-Battery Capacity 100% of the battery will provide 120 mins of monitoring; and 30 discharges at 360 joules.
ECG
ECG lead: I, II, III, AVR, AVL, AVF, V1-V6, GND S-T segment analysis: -2.0-2.0mv
Arrhythmia analysis: yes
Protection: With stand 4000vAc/50Hz voltage in isolation and work against electro surgical knife and defibrillation
Respiration
Measurement method: Thoracic impedance
NIBP
Working mode: Manual/Automatic/stat(5 minutes) Method: Automatic oscillation Cuff type: standard Adult; optional Pediatric/Neonatal
Temperature
Measurement range 20~45eC Resolution 0.1eC
Pulse Oximetry(spO2)
Display:spo2 value, pulse histogram, waveform, Pulse Range:0~99%for adult, pediatric and neonatal patients Probe: standard adult finger clip
Optional pediatric Y-type clip and neonatal wrap
Pulse range Pulse range: 0-255bpm
Standard parameter & configuration
ECG/ RESP,/SPO2,/TEMP,/PR,/NIBP/DEFIBRILLATOR
Optional parameter
Recorder